BE ORIGINATIVE NICHIRYO

ISO9001ISO14001

  • About NICHIRYO AMERICA
  • Environment
  • NICHIRYO GLOBAL SITE

MODEL 8100

FEATURES

  • 35 ways of dispensing in the volume range between 1 and 5000ul
  • Fast change of syringe and volume adjustment dial
  • Light and slender design permits easy handling without fatigue
  • Autoclavable syringe (except 3.0 ml and 60 ml
¦Dispensing volumes
Size of
syringe
0.06ml
(SG-V2)
0.6ml
(SG-S)
1.5ml
(SG-M)
3.0ml
(SG-N3)
6.0ml
(SG-L)
15.0ml
(SG-Y)
60.0ml
(SG-Z)
Maximum
Number of
Dispense
Volume
Adjust-
ment
Dial
1 1ul 10ul 25ul 50ul 100ul 250ul 1.0ml 58
2 2ul 20ul 50ul 100ul 200ul 500ul 2.0ml 28
3 3ul 30ul 75ul 150ul 300ul 750ul 3.0ml 18
4 4ul 40ul 100ul 200ul 400ul 1000ul 4.0ml 13
5 5ul 50ul 125ul 250ul 500ul 1250ul 5.0ml 10
¦Ordering information
Cat. No. Description Quantity
8100 One instrument with one each of 6.0ml,1.5ml and 0.6ml syringe. 1
SG-V2 0.06ml Syringe without tip 100
SG-S 0.6ml Syringe with a tip 100
SG-M 1.5ml Syringe with a tip 100
SG-MP 1.5ml Syringe without tip,but with a spacer in the nozzle
specially designed for a microtiter plate with low liquid pressure.
100
SG-N3 3.0ml Syringe without tip 100
SG-L 6.0ml Syringe without tip 100
SG-Y 15.0ml Syringe without tip 50
SG-Z 60.0ml Syringe without tip,but with one adaptor to mount
the syringe to the handle
50
8100S Stand to hold two dispensers with syringes 1
¦Accuracy and precision
Volume Accuracy Precision
SG-V2,1 ~ 5ul <±2.5% <1.5%~<1.0%
SG-S,10 ~ 50ul <±1.5% <1.0%~<0.8%
SG-M,25 ~ 125ul <±1.2% <0.9%~<0.6%
SG-N3,50 ~ 250ul <±1.0% <0.8%~<0.4%
SG-L,100 ~ 500ul <±0.8% <0.6%~<0.3%
SG-Y,250 ~ 1250ul <±0.8% <1.0%~<0.5%
SG-Z,1 ~ 5ml <±1.0% <4.0%~<2.5%

Above data of SG-S and SG-M are taken with the syringe having a Nichiryo pipette tip on the nozzle.